Multifunction control and evaluation of a 3D printed hand prosthesis with the Myo armband by hand amputees

Abstract

Upper limb amputations are highly impairing injuries that can substantially limit the quality of life of a person. The most advanced dexterous prosthetic hands have remarkable mechanical features. However, in most cases, the control systems are a simple extension of basic control protocols, making the use of the prosthesis not intuitive and sometimes complex. Furthermore, the cost of dexterous prosthetic hands is often prohibitive, especially for the pediatric population and developing countries. 3D printed hand prostheses can represent an opportunity for the future. Open 3D models are increasingly being released, even for dexterous prostheses that are capable of moving each finger individually and actively rotating the thumb. However, the usage and test of such devices by hand amputees (using electromyography and classification methods) is not well explored. The aim of this article is to investigate the usage of a cost-effective system composed of a 3D printed hand prosthesis and a low-cost myoelectric armband. Two subjects with transradial amputation were asked to wear a custom-made socket supporting the HANDi Hand and the Thalmic Labs Myo armband. Afterwards, the subjects were asked to control and use the prosthetic hand to grasp several objects by attempting to perform a set of different hand gestures. Both the HANDi Hand and the Myo armband performed well during the test, which is encouraging considering that the HANDi Hand was developed as a research platform. The results are promising and show the feasibility of the multifunction control of dexterous 3D printed hand prostheses based on low-cost setups. Factors as the level of the amputation, neuromuscular fatigue and mechanical limitations of the 3D printed hand prosthesis can influence the performance of the setup. Practical aspects such as usability and robustness will need to be addressed for successful application in daily life. A video of the tests can be found at the following link: https://youtu.be/iPSCAbd17Qw

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Cognolato, M., Atzori, M., Marchesini, C., Marangon, S., Faccio, D., Tiengo, C., Bassetto, F., Gassert, R., Petrone, N., Müller, H.. 2018-10-17. Multifunction control and evaluation of a 3D printed hand prosthesis with the Myo armband by hand amputees. https://doi.org/10.1101/445460

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related preprints

Impaired loading following denervation elicits dynamic, differential growth trajectories of musculoskeletal tissues

Physical forces dictate the function and maintenance of adult musculoskeletal tissues. Yet, the impact of biophysical cues on the growth and maturation of these tissues, especially dense connective tissues, remains unclear. Here, we investigated the effects of neonatal mechanical loading on early postnatal growth of the musculoskeletal tissues of the lower limb, with particular focus on the Achilles tendon. We performed sciatic nerve resection (SNR) on one-day-old mice to denervate the hamstrings and distal hindlimb muscles. Sciatic Nerve Resection resulted in severe and sustained alterations in gait, accompanied by an altered transcriptional landscape in distal musculoskeletal tissues, with marked reductions in bone and tendon growth and smaller, fibrotic hindlimb musculature. Sciatic nerve resection impaired cumulative growth in the Achilles tendon, shifting cell proliferation and matrix-driven growth trajectories and altering mechanotransduction gene programs toward inflammation and matrix catabolism with age. This work highlights the essential role of mechanical loading in driving mechanosignaling and postnatal growth trajectories in musculoskeletal tissues of the developing limb, with implications for overall musculoskeletal health in adulthood.

bioengineering↗

MiniMAP: An automatic pipeline for paediatric lower limb musculoskeletal model generation

The generation of paediatric musculoskeletal models often requires compromise between data availability, accuracy, and computation time. Statistical shape models allow for rapid generation of bone geometries for use in musculoskeletal modelling. An articulated shape model allows for movement at the joints during optimisation to better fit to experimental data. Here we present a methodology for generating custom paediatric OpenSim models using an articulated shape model of the lower limb generated from CT scans of over 300 children aged 4 to 18 years old. Bone geometries are predicted using demographic factors (age, height, mass, and sex), linear bone measurements (bone widths and lengths), and 12 skin surface landmarks (L/R ASIS, L/R PSIS, medial/lateral epicondyles, and medial/lateral malleoli). To account for skin padding (bone to skin surface distance) a random forest algorithm was developed for skin padding prediction for these 12 landmarks. Accuracy of the model prediction was assessed using a leave-one-out analysis to assess landmark errors, bone surface errors, and clinical bone measurement errors (linear measurements and angular values). Motion capture and MRI data were collected from two paediatric test cases to test the workflow on unseen data. Skin padding values varied widely in this population with average values from 3.4mm at the lateral malleolus to 25.9mm at the PSIS. Skin padding was best predicted using age, height, mass, ASIS width, epicondylar width, and malleolar width with an average RMSE of 4.68mm. The articulated shape model predicted accurate bone geometries from skin surface landmarks with average bone surface RMSEs of 2.64 {+/-} 0.99mm, 1.91 {+/-} 0.59mm, and 1.69 {+/-} 0.56mm in the pelvis, femora, and tibiae/fibulae respectively. In the two unseen test cases, average bone surface errors ranged from 1.64 {+/-} 1.2mm to 3.41 {+/-} 3.1mm. This study demonstrated the accuracy of "MiniMAP", for the rapid generation of custom paediatric musculoskeletal models.

bioengineering↗

A single-plane electrical impedance tomography framework for integrated liver and kidney assessment

Chronic liver and kidney diseases frequently coexist, but their assessment typically requires separate clinical workflows involving serum-based estimated glomerular filtration rate (eGFR) testing and liver elastography-based controlled attenuation parameter (CAP). Electrical impedance tomography (EIT) has recently shown promise for providing portable, non-invasive proxies for both reference measurements. However, existing multi-organ EIT assessment still requires different acquisition protocols for each organ. To simplify this procedure, we assessed the feasibility of a single-plane EIT framework for reconstructing organ-specific images and estimating liver and kidney health from one abdominal belt placement. Across diverse abdominal belt locations, opposite-plane deviation was significantly greater for eGFREIT estimates (2.92%; IQR, 0.46-12.53%) than for CAPEIT estimates (0.94%; IQR, 0.66-1.70%) (p < 0.01). For both organs, compared to images acquired at the conventional organ-specific planes, image correlation significantly decreased (both, p < 0.05) and normalized root mean squared error significantly increased (both, p < 0.05) at the furthest imaging plane, with these differences being numerically larger for the kidney. Together, these results supported the kidney plane as the preferred location for multi-organ assessment. Test-retest evaluation after belt removal and reapplication demonstrated that the correlation between two repeated CAPEIT estimates was just as strong at the kidney plane (R2 = 0.98, p < 0.001) as it was at the conventional liver plane (R2 = 0.97, p < 0.001). Finally, in an independent clinical cohort, kidney-plane CAPEIT estimates were significantly higher in participants with a high metabolic liver-risk profile (i.e. diabetes and high blood pressure) (p < 0.005). CAPEIT estimates also distinguished high and low metabolic risk groups with an AUC of 0.756. Together, these findings support the feasibility of a portable multi-organ EIT imaging framework capable of extracting complementary liver and kidney information from a single belt placement.

bioengineering↗