Accuracy of robotic coil positioning during transcranial magnetic stimulation
Abstract
ObjectiveRobotic positioning systems for transcranial magnetic stimulation (TMS) promise improved accuracy and stability of coil placement, but there is limited data on their performance. This text investigates the usability, accuracy, and limitations of robotic coil placement with a commercial system, ANT Neuro, in a TMS study.\n\nApproach21 subjects underwent a total of 79 TMS sessions corresponding to 160 hours under robotic coil control. Coil position and orientation were monitored concurrently through an additional neuronavigation system.\n\nMain ResultsRobot setup took on average 14.5 min. The robot achieved low position and orientation error with median 1.34 mm and 3.48{degrees}. The error increased over time at a rate of 0.4%/minute for both position and orientation.\n\nSignificanceAfter the elimination of several limitations, robotic TMS systems promise to substantially improve the accuracy and stability of manual coil position and orientation. Lack of pressure feedback and of manual adjustment of all coil degrees of freedom were limitations of this robotic system.
Source connections
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Goetz, S. M., Kozyrkov, I. C., Luber, B., Lisanby, S. H., Murphy, D. L. K., Grill, W. M., Peterchev, A. V.. 2019-05-19. Accuracy of robotic coil positioning during transcranial magnetic stimulation. https://doi.org/10.1101/641456