bioRxiv · 10.1101/606418
Sound comparison of seven TMS coils at matched simulation strength
Abstract
BackgroundAccurate data on the sound emitted by transcranial magnetic stimulation (TMS) coils is lacking.\n\nMethodsWe recorded the sound waveforms of seven coils with high bandwidth. We estimated the neural stimulation strength by measuring the induced electric field and applying a strength-duration model to account for different waveforms.\n\nResultsAcross coils, at maximum stimulator output and 25 cm distance, the sound pressure level (SPL) was 98-125 dB(Z) per pulse and 75-97 dB(A) for a 15 Hz pulse train. At 5 cm distance, these values were estimated to increase to 112-139 dB(Z) and 89-111 dB(A), respectively.\n\nConclusionsThe coils sound was below, but near, relevant exposure limits for operators and may exceed some limits for the subject. Exposure standards may inadequately capture some risks to hearing. For persons near operating TMS coils we recommend hearing protection, and we consider it essential for the TMS subject.\n\nHighlights O_LICoil click varies by over 20 dB(Z) between TMS coils at matched stimulation strength.\nC_LIO_LIClose to TMS coil, sound pressure level may reach nearly 140 dB(Z).\nC_LIO_LIFor rTMS, the continuous sound level can exceed 110 dB(A).\nC_LIO_LIHearing protection is recommended during TMS, especially for the subject.\nC_LI
Source connections
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Koponen, L. M., Goetz, S. M., Peterchev, A. V.. 2019-04-12. Sound comparison of seven TMS coils at matched simulation strength. https://doi.org/10.1101/606418
Cite the original work for its findings. Save a collection to share your selection of sources.