bioRxiv Science⌕ Search

bioRxiv · 10.1101/2021.03.26.437212

Optimized hip-knee-ankle exoskeleton assistance at a range of walking speeds

Abstract

BackgroundEffective autonomous exoskeletons will need to be useful at a variety of walking speeds, but we do not know how optimal exoskeleton assistance should change with speed. Optimal exoskeleton assistance may increase with speed similar to biological torque changes or a well-tuned assistance profile may be effective at a variety of speeds. MethodsWe optimized hip-knee-ankle exoskeleton assistance to reduce metabolic cost for three participants walking at 1.0 m/s, 1.25 m/s and 1.5 m/s. We measured metabolic cost, muscle activity, exoskeleton assistance and kinematics. We performed two tailed paired t-tests to determine significance. ResultsExoskeleton assistance reduced the metabolic cost of walking compared to wearing the exoskeleton with no torque applied by 26%, 47% and 50% at 1.0, 1.25 and 1.5 m/s, respectively. For all three participants, optimized exoskeleton ankle torque was the smallest for slow walking, while hip and knee torque changed slightly with speed in ways that varied across participants. Total applied positive power increased with speed for all three participants, largely due to increased joint velocities, which consistently increased with speed. ConclusionsExoskeleton assistance is effective at a range of speeds and is most effective at medium and fast walking speeds. Exoskeleton assistance was less effective for slow walking, which may explain the limited success in reducing metabolic cost for patient populations through exoskeleton assistance. Exoskeleton designers may have more success when targeting activities and groups with faster walking speeds. Speed-related changes in optimized exoskeleton assistance varied by participant, indicating either the benefit of participant-specific tuning or that a wide variety of torque profiles are similarly effective.

Source connections

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Bryan, G. M., Franks, P. W., Song, S., Voloshina, A. S., Reyes, R., O'Donovan, M. P., Gregorczyk, K. N., Collins, S. H.. 2021-03-28. Optimized hip-knee-ankle exoskeleton assistance at a range of walking speeds. https://doi.org/10.1101/2021.03.26.437212

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related preprints

Preserving Nanotherapy-Induced Allogeneic Murine Islet Graft Tolerance During Distal Methicillin-Resistant Staphylococcus aureus Skin Infection

Eliminating lifelong immunosuppression can make islet transplantation a durable alternative to insulin therapy for type 1 diabetes. As short-course tolerogenic regimens are developed to replace lifelong immunosuppression, it is critical to test their robustness against perturbations such as hospital-acquired infections, which transplant recipients routinely encounter. Herein, we investigated whether rapamycin-loaded polymersomes (rPS), a subcutaneous nanotherapy that induces long-term murine islet allograft survival, can preserve host defenses in a Methicillin-resistant Staphylococcus aureus (MRSA) skin infection, and whether the resulting immune response from this distal infection relative to graft location erodes rPS-induced tolerance. rPS maintained neutrophil antibacterial activity despite inhibiting oxidative mechanisms in vitro, and rPS-treated mice demonstrated normal recovery following MRSA challenge. However, MRSA skin infection caused a decline of ~43% in recipient survival for rPS-treated mice, despite surviving recipients retaining glucose-stimulated insulin responsiveness. This erosion of allograft acceptance coincided with a burst of IL-6 within 24h of the infection followed by elevated TNF-[a] and neutrophilia at 13 days post-infection. Combining rPS with a single dose of methylprednisolone reduced IL-6 and TNF- without altering neutrophilia, nearly doubling survival among infected recipients to 80%. Our findings demonstrate that partial dampening of inflammatory insults can restore superior islet allograft acceptance following perioperative infection.

bioengineering↗

Extensional Flow Microfluidics Reveal Size-Dependent Core-Shell Mechanical Heterogeneity in Pancreatic Tumor Spheroids

Understanding the mechanical behavior of cancer cell aggregates is important for investigating metastatic potential and developing effective therapeutics. Tumor cell clusters are major contributors to metastasis and can undergo extreme deformation to migrate through capillaries smaller than the cluster size, suggesting complex multicellular mechanics. Despite their clinical significance, the mechanics of multicellular spheroids remain underexplored. In this study, the mechanical behavior of spheroids formed from the pancreatic cancer cell line PANC-1 was investigated using a cross-slot microfluidic device. Suspended spheroids from two size ranges were stretched in non-contact extensional flow (small spheroids mostly 40 < D0 < 75 m, large spheroids mostly 90 < D0 < 145 m). Digital image correlation analyses of intra-spheroid strains revealed strong deformation differences for small and large spheroids while analyses based on 2D projected contours of the spheroid shape did not show trends. Smaller spheroids deformed relatively homogeneously and exhibited significant overall stretching. In contrast, larger spheroids displayed heterogeneous deformation characterized by a stiff central core and a deformable outer shell. To investigate the origin of these mechanical differences, cellular morphology and the expression of stiffness-associated proteins, including vimentin and E-cadherin, were assessed from confocal microscopy. Larger spheroids display significantly lower nuclear sphericity than small spheroids, indicating more cellular packing with increasing spheroid size. Additionally, vimentin expression was concentrated in the outer shell of large spheroids, whereas small spheroids exhibit vimentin expression throughout the aggregate. Thus individual cell morphology and cytoskeletal networks seemed to contribute to the core-shell biomechanical behavior of the large spheroids and lack thereof for small spheroids. Taken together, this study suggests that as spheroids grow, they transition from an irregular and disorganized cluster at low cell numbers to a coordinated and hierarchical structure at sufficient cell numbers that corresponds to diameters {gtrsim} 100 m.

bioengineering↗

Material context and TGF-β blockade timing determine retained matrix in patient-derived bioprinted keratoconus stroma

Stromal repair requires newly formed matrix to persist while the contractile response to TGF-{beta} is controlled. We developed patient-derived bioprinted gelatin methacryloyl (GelMA) stroma to connect material-dependent cell responses with matrix retention after timed receptor blockade. In constructs from six keratoconus (KC) and six normal donors, KC cells retained disease-associated transcription while secreting procollagen I and lumican at normal-donor levels per unit DNA. Comparisons with two-dimensional culture and matched cast GelMA showed that the three-dimensional environment increased matrix induction and lowered resting contractile expression. Printing added a stromal identity response that persisted as retained lumican. Delayed blockade after a TGF-{beta}1 pulse preserved matrix transcription and late secretion. Eleven days after withdrawal, this finite course left more native collagen and lower contractile-gene expression than vehicle in 9/12 donors. Across the tested blockade regimens, at least one regimen achieved both outcomes at the final endpoint in 11/12 donors, including every KC donor. Culture format changed the preferred regimen. Choosing finite blockade by retained collagen instead of secretion changed the schedule in half the donors; the selected arms contained 22.2% more collagen. Retained collagen and contractile expression therefore provide joint endpoints for designing when TGF-{beta}-pathway courses, including losartan and TGF-{beta}3 treatment for keratoconic and scarred corneas, should start and stop.

bioengineering↗