Real-time ultrafast holography of focused ultrasound wave fields using an ultrasound scanner

Abstract

Acoustic pressure field measurements are required for marketing clearance of medical ultrasound instruments, but the gold standard method, hydrophone scanning, is time consuming. The objective of this work was to develop an ultrafast holographic method for real-time imaging of focused ultrasound fields. The technique is based on parallel acquisition with a sensor array, and FUS field reconstruction based on time reversal assuming a homogeneous lossless medium. The geometric characteristics of the sensor to enable accurate reconstruction were evaluated through numerical modelling. A practical example involving a commercially-available 128-element linear-array ultrasound imaging probe is presented. Finally, preliminary experiments were conducted on in-vitro soft tissues to validate the concept of simultaneous dual-mode imaging, combining conventional B-mode tissue imaging with FUS field holography. In modelling of ideal sensor characteristics (2D matrix; intercepted energy > 99%), parallel holography allowed reconstructing wave fields with excellent agreement with the Rayleigh Integral reference (errors << 0.01%). Successfully experimental demonstration of real-time ultrafast wave field holography was achieved with the commercial imaging probe (7-10% error in the -6dB focus vs hydrophone measurements, acquisition time: 100 microseconds, display rate: 15 frames/s), despite a suboptimal intercepted energy (44%). Real-time dual-mode imaging of continuous wave fields allowed targeting localized tissues regions, while holographic movies highlighted the propagation of transient waves. Overall, real-time ultrafast holography offers a practical way to perform metrology and guidance of ultrasonic devices, by visualising waves fields typically used for ultrasound imaging, or those used during thermal or non-thermal ultrasound therapy.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Matsui, K., Chavrier, F., Azuma, T., Sakuma, I., N'Djin, A., Souchon, R.. 2026-10-16. Real-time ultrafast holography of focused ultrasound wave fields using an ultrasound scanner. https://doi.org/10.64898/2026.10.15.759327

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related preprints

Development of a Low-cost Extrusion Multi-nozzle 3D Bioprinter for Complex Constructs

Extrusion-based multi-material bioprinting is a promising approach for fabricating complex tissue constructs, but current systems often require expensive hardware or support only one or a few numbers of bioinks. In this study, a low-cost, compact, and versatile multi-nozzle bioprinter (MNB) system was developed for pneumatic extrusion of up to 10 different inks using a rotational head. The system was integrated with a commercial 3D printer platform and controlled using custom G-code commands. Backlash compensation and nozzle-specific calibrations were experimentally determined to minimize positioning errors and airflow misalignment. The developed system showed stable repeatability in both clockwise and counterclockwise movements. Lines and various structures were successfully printed with different nozzle-switching directions and printing sequences. Representative structures including a diamond pattern, a multilayer scaffold, a pyramid structure, and a kidney-like construct were successfully fabricated. These results demonstrate that the proposed MNB system can produce complex 2D and 3D multi-material structures with good positional accuracy and repeatability.

bioengineering↗

FluOpti: an open-source system for programmable whole-plate optogenetics and multi-fluorescence imaging

Bacteria are able to self-organize into complex spatial communities, such as colonies, biofilms and swarms. Probing cells and gene networks with dynamic stimuli while preserving their spatial organization is therefore essential for understanding how bacterial communities organize across spatial and temporal scales. Optogenetics offers a powerful approach for such a task given the precise control of stimulus delivery without the perturbations normally associated with the addition of liquid inducers. However, most optogenetics devices have been developed to work with liquid cell cultures or to deliver a spatial pattern without the concurrent recording of gene responses and colony growth. Here, we describe FluOpti, an open source, low-cost device for controlling gene expression, through programmable light inputs, and recording multi-fluorescence responses and colony growth across whole petri dishes. White light LEDs are used for brightfield imaging while green and red LEDs are used as optogenetic inputs to activate and inactivate target genes via the CcaS/R two component system in E. coli. Low cost blue and orange acrylic are used as excitation and emission filters for fluorescence imaging, respectively. We developed custom software to analyze colony growth and optogenetic responses from 72-hour time-lapse experiments. A mathematical model was developed to predict gene expression levels in response to green-to-red light ratio stimuli. We use FluOpti to reveal a spatially structured response to optogenetic stimuli, with gene activation rates differing in cells located at the center and edge of colonies. to validate its use at a whole petri dish scale, FluOpti was used to depict how Voronoi territories can act as predictors of colony size, in congruence with previous studies. Together, these results demonstrate the utility of the device for studying gene networks at multiple spatio-temporal scales, from individual colonies to the whole petri dish. Performing optogenetics studies in whole petri dishes while preserving spatial arrangements could offer valuable insights into the spatial biology of gene networks in microbial populations.

bioengineering↗

Denoising broadband CARS hyperspectral images on variance-stabilized data: a unified benchmark

Broadband coherent anti-Stokes Raman scattering (BCARS) enables label-free hyperspectral chemical imaging, but weak signal components are often obscured by heteroscedastic noise. Here, we develop a physics-based variance stabilization transform and benchmark six hyperspectral denoising methods to identify effective strategies for BCARS noise reduction. The transform yields nearly intensity-independent noise variance while preserving the dispersive Raman response. Within this framework, we compare singular value decomposition, BM4D, Noise2Noise, deep image prior, self-supervised diffusion, and a cross-covariance-based singular value decomposition method (CCV) introduced here, using a shared phase-retrieval pipeline. Evaluation includes five simulated hyperspectral cubes with exact ground truth and experimental measurements of glycerol, polystyrene beads, and Caenorhabditis elegans. In simulations, variance stabilization reduces the ratio of maximum to minimum channel noise variance from approximately 72 to 1.2. Deep image prior and self-supervised diffusion achieve the highest mean foreground peak signal-to-noise ratios. CCV matches the mean foreground structural similarity of deep image prior with substantially shorter runtime and no per-cube rank tuning. Experimental comparisons reveal sample-dependent performance. These results establish physics-based variance stabilization as a common framework for adapting and evaluating hyperspectral denoisers for BCARS, with CCV offering a practical balance of reconstruction fidelity, computational speed, and automated operation.

bioengineering↗