bioRxiv · 10.64898/2026.09.30.755689
BrightEyes-Scope: A Reference Implementation for Image Scanning Microscopy with SPAD Array Detection
Abstract
Image scanning microscopy (ISM) is emerging as the natural evolution of confocal laser-scanning microscopy, combining optical sectioning and a two-fold improvement in lateral resolution with high signal-to-noise ratio. Single-photon avalanche diode (SPAD) array detectors enable an effective implementation of ISM by recording a micro-image at each scan position with single-photon sensitivity, negligible readout noise, and without the fixed frame-readout constraints of conventional cameras. Nevertheless, SPAD array-based ISM remains largely confined to laboratories with expertise in microscope development. Here, we present BrightEyes-Scope, a reference implementation of a SPAD array-based ISM system. We provide step-by-step guidelines for optical design, system construction, alignment, calibration, data acquisition, and image reconstruction. Presented as a complete ISM microscope, BrightEyes-Scope also serves as a reference for upgrading existing laser-scanning microscopes. We anticipate that it will accelerate the adoption of ISM for quantitative super-resolution imaging.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
LE, L., Zunino, A., Slenders, E., Bucci, A., Donato, M., Vicidomini, G.. 2026-10-02. BrightEyes-Scope: A Reference Implementation for Image Scanning Microscopy with SPAD Array Detection. https://doi.org/10.64898/2026.09.30.755689
Cite the original work for its findings. Save a collection to share your selection of sources.