Characterization of real B0 shim fields generated by higher order B0 shim systems of whole body human 3T and 7T MRI systems

Abstract

PurposeEnsuring optimal homogeneity of the static magnetic field (B0) is critically important for significantly enhancing the image quality, spectral resolution, and overall diagnostic accuracy of magnetic resonance imaging (MRI) and magnetic resonance spectroscopy/spectroscopic imaging (MRS/MRSI), especially at high and ultra-high field strengths. Achieving this goal relies on the deployment of advanced, higher-order shim hardware, which is indispensable for effective and precise B0 shimming. MethodsIn this study, we acquired B0 field maps to characterize the performance of each shim channel of vendor-provided B0 shim hardware integrated into whole body human 7T MRI system. To rigorously evaluate and compare the B0 field homogeneity achievable using these devices, we utilized spherical harmonic expansions to fit the measured B0 distributions for each shim channel in each of the investigated built-in B0 shim systems. This methodology enabled a systematic assessment of shim field "purity" and the effectiveness of different B0 shim configurations. ResultsWhile our analysis confirmed that the linear-order shim terms provided by the gradient coils maintained high purity, certain second-order shim terms and all third-order B0 shim coils of all vendors 7T whole body human MRI systems exhibit a high level of impurity and more specifically introduce substantial linear field components. These unintended linear contributions interfere with the linear shim fields provided by the gradient coils negating their corrective influence. As a result, the 3rd order B0 shim coils and some 2nd order B0 shim coil elements compromised the overall effectiveness of the B0 shimming process rather than improving it. Although pure higher order B0 shim fields were previously shown to substantially improve B0 field inhomogeneity, the vendor implemented versions of higher order B0 shim systems in current 7T whole-body human MRI scanners cannot achieve the desired level of uniformity and partly perform worse than first order B0 shimming. Substantial deviations between 2nd order and especially 3rd order real shim fields versus ideal shim fields demonstrate the necessity of real shim field calibration and call for improvement of future 2nd and 3rd order B0 shim coil design for human whole-body MRI scanners. Our findings offer valuable guidance for optimizing B0 shimming strategies at high and ultra-high field strengths, ultimately enhancing image quality. ConclusionRecent evaluations have demonstrated that certain second- and third-order shim coils, as implemented in current commercial 7T human whole-body MRI systems, introduce unintended lower-order (linear) field components. These field "impurities" substantially deviate from the desired higher-order spatial harmonics, thereby diminishing the orthogonality of the higher order B0 shim system. This in turn largely compromises the effectiveness of higher-order B0 shimming at whole-body 7T MRI scanners. These findings emphasize the necessity of conducting accurate shim field calibration to best utilize the current 7T B0 shim system configurations on one hand. On the other hand, improving the purity of these commercial B0 shim systems in the design phase to ensure that the higher-order terms accurately match their intended spherical harmonic profiles, particularly for the third-order shim terms--would largely enhance the achievable B0 field homogeneity at 7T. This improvement has directly clinically significant outcomes.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Jani, M., Su, S., Gast, L., Nagel, A., Henning, A.. 2025-08-27. Characterization of real B0 shim fields generated by higher order B0 shim systems of whole body human 3T and 7T MRI systems. https://doi.org/10.1101/2025.08.26.672172

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related preprints

Wall stiffening is a primary contributor to motility loss in Crohn's disease: an electromechanical modeling study

Fibrotic strictures are among the most disabling complications of Crohn's disease, permanently narrowing the bowel and impairing motility, yet no approved therapy reverses them. Chronic inflammation alters pacemaker-network coupling, smooth-muscle excitability, and calcium-dependent contractility, while fibrosis thickens the bowel wall, narrows the lumen, and changes tissue mechanics. The relative contributions of these coupled electrical, contractile, and structural alterations to motility loss remain unclear. To address this gap, we develop an integrated electromechanical finite-element framework for fibrostenosing Crohn's disease that couples a fibrosis-driven growth model with a FitzHugh-Nagumo electromechanical model. A full-factorial 25 design of experiments is used to quantify the relative effects of electrical diffusivity, excitation threshold, peak active stress, wall stiffness, and hypertrophic remodeling on cyclic lumen-volume deformation. Motility is quantified by the standard deviation of lumen volume over one contraction cycle. Within the parameter ranges examined, increased wall stiffness emerged as the dominant contributor to motility loss, followed by impaired smooth-muscle contractility. Changes in excitation threshold, hypertrophic remodeling, and electrical diffusivity produced substantially smaller effects. Pairwise interactions were small relative to the dominant main effects, indicating that the mechanisms contributed largely through their individual effects. Our findings suggest that limiting wall stiffening while preserving smooth-muscle contractile function may provide a therapeutic strategy for maintaining intestinal motility in fibrostenosing Crohn's disease.

bioengineering↗

Lactate Receptor Activation Alleviates Senescence and Preserves Homeostasis of Aged Arteries

Arteries are among the first tissues to exhibit age-related dysfunction, yet the metabolic mechanisms driving vascular senescence remain poorly understood. Here, analysis of human aortic transcriptomic data identified HCAR1, encoding the lactate receptor GPR81, as one of the genes most significantly downregulated with age. We therefore investigated whether age-associated loss of GPR81 contributes to cellular senescence within the vessel wall. Senescent human endothelial cells and vascular smooth muscle cells accumulated neutral and oxidized lipids and exhibited increased labile iron and ferroptosis. Silencing GPR81 in early-passage cells recapitulated this metabolic phenotype together with multiple hallmarks of cellular senescence. Moreover, endothelial-specific deletion of GPR81 in young mice was sufficient to induce senescent cell accumulation, impaired lipid homeostasis, endothelial dysfunction, and elastin disorganization. Conversely, pharmacological activation of GPR81 with the agonist CHBA restored fatty acid metabolism, promoted glycolytic reprogramming, and attenuated ferroptotic stress and senescence-associated phenotypes. In lamin A knock-in (LAKI) progeroid mice, CHBA reduced arterial lipid accumulation and cellular senescence, shifted vascular cell composition toward a youthful state, improved endothelial integrity, and restored extracellular matrix homeostasis. Together, these findings identify age-associated loss of GPR81 as a driver of vascular metabolic dysfunction and cellular senescence and establish pharmacological GPR81 activation as a promising therapeutic strategy for preserving vascular homeostasis and mitigating age-associated cardiovascular disease.

bioengineering↗

Targeted and bilateral blood flow monitoring in middle cerebral artery using diffuse correlation spectroscopy

Objective: To develop and validate a dual-probe Diffuse Correlation Spectroscopy (DCS) system for non-invasive and simultaneous, monitoring of cerebral blood flow (CBF) in the bilateral Middle Cerebral Artery (MCA) territories, and expanding the utility of conventional DCS limited to cortical-volume-based CBF measurements to vessel-specific cerebral perfusion monitoring. Methods: A dual-probe DCS system was designed for non-invasive monitoring of MCA-specific perfusion. Probe placement and protocol optimization study has been performed using anatomical landmarks, motor and speech activation tasks in healthy volunteers. System stability and repeatability were further evaluated in a pilot cohort of 30 healthy (age, 25{+/-}7 years) participants using optimized probe position and protocol. A bilateral MCA ischemic Lacunar Infract stroke case report also validated the feasibility of the system in clinical settings. Results: Measurements demonstrated superior sensitivity towards MCA-territory perfusion at targeted probe locations compared to off-MCA positions. In pilot cohort, significant increase of 30.34 {+/-} 21.56% and 36.48 {+/-} 21.22% in rCBF corresponding to hand squeeze and speech task respectively showed reproducible physiological responsiveness of the system (p<0.001). Measurement done on a patient with bilateral MCA ischemic Lacunar Infract stroke showed a significant change of 30% during speech for both the MCAs but no significant change is observed for hand squeeze tasks. Conclusion: The custom built dual-probe DCS system enables non-invasive, operator-independent, targeted and continuous monitoring of rCBF within bilateral MCA territories. Significance: This approach enables the potential use of DCS system for bilateral and vessel-specific monitoring of cerebral perfusion in the MCA territories.

bioengineering↗