In vivo validation of an in situ calibration bead as a reference for backscatter coefficient calculation

Abstract

ObjectivesThe study aims to assess the capability of Quantitative Ul-trasound (QUS) based on the backscatter coefficient (BSC) for classifying disease states, such as breast cancer response to neoadjuvant chemotherapy and quantifying fatty liver disease. We evaluate the effectiveness of an in situ titanium (Ti) bead as a reference target in calibrating the system and mitigating attenuation and transmission loss effects on BSC estimation. MethodsTraditional BSC estimation methods require external references for calibration, which do not account for ultrasound attenuation or transmis-sion losses through tissues. To address this issue, we use an in situ titanium (Ti) bead as a reference target, because it can be used to calibrate the system and mitigate the attenuation and transmission loss effects on estimation of the BSC. The capabilities of the in situ calibration approach were assessed by quantifying consistency of BSC estimates from rabbit mammary tumors (N = 21). Specifically, mammary tumors were grown in rabbits and when a tumor reached 1 cm or greater in size, a 2-mm Ti bead was implanted into the tumor as a radiological marker and a calibration source for ultrasound. Three days later, the tumors were scanned with a L-14/5 38 array transducer connected to a SonixOne scanner with and without a slab of pork belly placed on top of the tumors. The pork belly acted as an additional source of attenu-ation and transmission loss. QUS parameters, specifically effective scatterer diameter (ESD) and effective acoustic concentration (EAC), were calculated using calibration spectra from both an external reference phantom and the Ti bead. ResultsFor ESD estimation, the 95% confidence interval between measure-ments with and without the pork belly layer was (6.0,27.4) using the in situ bead and (114, 135.1) with the external reference phantom. For EAC esti-mation, the 95% confidence interval were (-8.1, 0.5) for the bead and (-41.5, -32.2) for the phantom. These results indicate that the in situ bead method shows reduced bias in QUS estimates due to intervening tissue losses. ConclusionsThe use of an in situ Ti bead as a radiological marker not only serves its traditional role but also effectively acts as a calibration target for QUS methods. This approach accounts for attenuation and transmission losses in tissue, resulting in more accurate QUS estimates and offering a promising method for enhanced disease state classification in clinical settings.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Zhao, Y., Oelze, M., Park, T. H., Miller, R. J., Czarnota, G.. 2024-02-09. In vivo validation of an in situ calibration bead as a reference for backscatter coefficient calculation. https://doi.org/10.1101/2024.02.07.579320

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related preprints

Comparative study of chlorophyll measurement in Physcomitrium patens moss using a conventional microscope adapted for combined 2D+1D imaging and spectral analysis

Imaging spectroscopy often requires expensive and complex equipment. Here we show a simple procedure for attaching a standard miniature fiber spectrometer to a conventional microscope, allowing easy integration of 2D imaging with 1D high-resolution spectral measurements. This combination provides much of the benefit of a full imaging spectrometer without the large equipment investment, and we provide instructions for modifying microscopes to this setup and the present measurements of living cells that demonstrate their performance. Using this setup, we compare the quantitative measurement of chlorophyll concentration in Physcomitrium patens moss using color imaging and spectral sampling.

bioengineering↗

De novo designed single-domain antibodies protect against lethal cobra venom neurotoxicity in vivo

Generative protein design can now rapidly produce de novo binders with high affinity and functional activity against a wide range of targets, including lethal snake venom toxins. However, so far most reported successes rely on new-to-nature scaffolds with limited therapeutic precedent. Single-domain antibodies (VHHs) offer a clinically validated alternative scaffold that can bind and neutralize long-chain -neurotoxins, which are some of the most lethal components in snake venoms. Here we compare three recently established de novo design models with VHH-design capabilities (Germinal, RFantibody, and BoltzGen) for their ability to generate VHHs against the neurotoxin -cobratoxin from the monocled cobra (Naja kaouthia). Using standardized model inputs and evaluation criteria based on AlphaFold3 interface confidence (ipTM) and RMSD self-consistency, we find that Germinal was the only method to generate designs passing stringent in silico criteria for experimental testing. We therefore performed a larger Germinal design campaign employing three different VHH frameworks and experimentally validated 46 designs in vitro. Of these, 42 expressed as soluble proteins and we identified four binding hits derived from two of the three tested frameworks. Of the four binders, two lead candidates were further characterized and demonstrated high affinity (KDs of 4.1 nM and 10.8 nM), monomeric behavior and low polyreactivity, indicating favorable biophysical and developability properties, as well as functional toxin neutralization in vitro. To assess their therapeutic potential we investigated their ability to protect against -cobratoxin toxicity in vivo. Both candidates fully protected mice after -cobratoxin challenge, with 100% survival compared to a lethal control. One candidate also retained notable neutralization capacity against whole venom of Naja kaouthia with a survival of 56%, while the other protected 22% when tested in a rescue setting. Together, we demonstrate that de novo VHH design can generate high affinity single-domain antibodies with in vivo protection against lethal cobra venom neurotoxicity, and provide practical insights into method- and framework-dependent performance.

bioengineering↗

Simple Feedback for Complex Movement: Capturing Whole-Limb Reorganization during Single-IMU Gait Retraining

Clinical gait retraining typically relies on multi-sensor arrays and high-dimensional feedback displays, imposing setup and interpretation burdens that limit routine clinical deployment. We developed a single-IMU visual biofeedback system that delivers real-time feedback of Lower Limb Trajectory Error (LLTE), a composite kinematic error metric integrating knee position and shank angle across the stance phase. Twenty able-bodied adults walked on a treadmill under two visual biofeedback targets (flexed-knee, extended-knee) while receiving either corrected (n=10) or uncorrected (n=8) feedback, where the correction accounted for limb orientation at initial contact. LLTE and stance-phase knee kinematics adapted consistently under the flexed-knee target for both feedback groups, with feedback formulation moderating the temporal trajectory of change. Adaptation toward the extended-knee target was limited, likely because participants were already operating near terminal knee extension and because the scalar error metric provided limited directional information for correction. Ankle range of motion (ROM) changed significantly across the stance phase under both target conditions, while hip ROM did not. Multiscale multivariate sample entropy (MSMVSE) increased monotonically with time scale across all conditions, with no statistically distinguishable difference between corrected and uncorrected feedback. These results suggest that single-IMU LLTE biofeedback can modify gait mechanics and that adaptation was expressed across multiple lower-limb segments rather than through changes at a single joint.

bioengineering↗