bioRxiv · 10.1101/2023.02.12.528160
Overcoming resolution loss due to thermal magnetic field fluctuations from phase plates in transmission electron microscopy
Abstract
We identify thermal magnetic field fluctuations, caused by thermal electron motion ("Johnson noise") in electrically conductive materials, as a potential resolution limit in transmission electron microscopy with a phase plate. Specifically, resolution loss can occur if the electron diffraction pattern is magnified to extend phase contrast to lower spatial frequencies, and if conductive materials are placed too close to the electron beam. While our initial implementation of a laser phase plate (LPP) was significantly affected by these factors, a redesign eliminated the problem and brought the performance close to the expected level. The resolution now appears to be limited by residual Johnson noise arising from the electron beam liner tube in the region of the LPP, together with the chromatic aberration of the relay optics. These two factors can be addressed during future development of the LPP.
Source connections
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Axelrod, J. J., Petrov, P. N., Zhang, J. T., Remis, J., Buijsse, B., Glaeser, R., Mueller, H.. 2023-02-13. Overcoming resolution loss due to thermal magnetic field fluctuations from phase plates in transmission electron microscopy. https://doi.org/10.1101/2023.02.12.528160