Effect of field strength on RF power deposition near conductive leads: A simulation study of SAR in DBS lead models during MRI at 1.5 T - 10.5 T

Abstract

BackgroundSince the advent of magnetic resonance imaging (MRI) nearly four decades ago, there has been a quest for ever-higher magnetic field strengths. Strong incentives exist to do so, as increasing the magnetic field strength increases the signal-to-noise ratio of images. However, ensuring patient safety becomes more challenging at high and ultrahigh field MRI (i.e., [≥]3 T) compared to lower fields. The problem is exacerbated for patients with conductive implants, such as those with deep brain stimulation (DBS) devices, as excessive local heating can occur around implanted lead tips. Despite extensive effort to assess radio frequency (RF) heating of implants during MRI at 1.5 T, a comparative study that systematically examines the effects of field strength and various exposure limits on RF heating is missing. PurposeThis study aims to perform numerical simulations that systematically compare RF power deposition near DBS lead models during MRI at common clinical and ultra-high field strengths, namely 1.5, 3, 7, and 10.5 T. Furthermore, we assess the effects of different exposure constraints on RF power deposition by imposing limits on either the B1+ or global head specific absorption rate (SAR) as these two exposure limits commonly appear in MRI guidelines. MethodsWe created 33 unique DBS lead models based on postoperative computed tomography (CT) images of patients with implanted DBS devices and performed electromagnetic simulations to evaluate the SAR of RF energy in the tissue surrounding lead tips during RF exposure at frequencies ranging from 64 MHz (1.5 T) to 447 MHz (10.5 T). The RF exposure was implemented via realistic MRI RF coil models created based on physical prototypes built in our institutions. We systematically examined the distribution of local SAR at different frequencies with the input coil power adjusted to either limit the B1+ or the global head SAR. ResultsThe MRI RF coils at higher resonant frequencies generated lower SARs around the lead tips when the global head SAR was constrained. The trend was reversed when the constraint was imposed on B1+. ConclusionAt higher static fields, MRI is not necessarily more dangerous than at lower fields for patients with conductive leads. Specifically, when a conservative safety criterion, such as constraints on the global SAR, is imposed, coils at a higher resonant frequency tend to generate a lower local SAR around implanted leads due to the decreased B1+ and, by proxy, E field levels.

Source connections

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Kazemivalipour, E., Sadeghi-Tarakameh, A., Keil, B., Eryaman, Y., Atalar, E., Rad, L. G.. 2022-12-15. Effect of field strength on RF power deposition near conductive leads: A simulation study of SAR in DBS lead models during MRI at 1.5 T - 10.5 T. https://doi.org/10.1101/2022.12.13.520303

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related preprints

A Minimally Perturbative DARPin Probe for Quantitative Fluorescence Imaging of the Human TCR-CD3 Complex

Fluorescence microscopy is a powerful tool for dissecting the molecular mechanisms of T-cell antigen recognition in living cells, but its quantitative insight critically depends on non-perturbative, high-quality probes. Here, we repurpose a small (~15 kDa) CD3epsilon-binding DARPin (designed ankyrin repeat proteins) to a fluorescent label for T-cell receptor (TCR)/CD3 complexes on primary human CD8+ T-cells, with the aim of generating a powerful tool for quantitative analysis, single-molecule tracking, and advanced imaging of TCR dynamics. We show that the DARPin binds CD3{varepsilon} with high affinity and selectivity and using single molecule tracking and brightness analysis, we characterize the TCR-CD3 diffusion behavior and show that the DARPin binds to both CD3epsilon; subunits. Importantly, labeling preserves antigen sensitivity: on supported lipid bilayers presenting cognate pMHC, T-cells remain responsive, assemble synapses, form TCR microclusters, and initiate signaling similar to unlabeled controls. We further demonstrate compatibility with lattice light-sheet microscopy for volumetric imaging of T-cell - APC interactions in living cells. Together, these results establish DARPins as versatile, minimally perturbative probes for high resolution, quantitative studies of T cell synapse organization and signaling.

biophysics↗

Monitoring intramolecular dynamics across two regions of the mouse prion protein during misfolding and oligomerization using fluorescence correlation spectroscopy

It is important to determine whether native state dynamics drive the misfolding and oligomerization of the prion protein, which are important events in prion disease, and how they are modulated by conformational conversion. Native (N) mouse prion protein (moPrP) is known to form small (OS) and large (OL) oligomers rich in {beta}-sheet, and in this study, photoinduced electron transfer-fluorescence correlation spectroscopy (PET-FCS) has been used to characterize intramolecular dynamics within individual monomeric units in both isolated OS and OL, as well as the diffusion properties of the oligomers. It is estimated that OS and OL comprise of about 15 and 55 monomeric units, respectively. Microsecond dynamics at each of the two regions that are the 1-3 and 2-3 interfaces of native protein are distinct in N, OS and OL, although they occur on very similar timescales. Analysis of the evolution of the distribution of diffusion times, determined using the maximum entropy method, indicates heterogeneity in the oligomerization reaction. Analysis of the change in the fluctuations which occur in two different timescales in the native state ensemble shows that they are damped more at the erstwhile 1-3 interface than the erstwhile 2-3 interface. The difference in the extent of damping at the erstwhile 1-3 and 2-3 interfaces can be explained on the basis of the structural changes known to occur across each region. The changes in dynamics occur concurrently in both regions, indicating that the structural changes accompanying conformational conversion also occur simultaneously during the oligomerization of moPrP.

biophysics↗

Combining CHARMM36m with OPC water improves accuracy

Atomistic simulations of intrinsically disordered proteins (IDPs) are notoriously sensitive to force field inaccuracies, either regarding the protein or the water model, yielding inaccurate observables such as compactness, secondary structure propensities, or kinetics. The currently most widely used IDP force fields are Amber99sb-disp (A99disp) and Charmm36m (C36m). A99disp includes a new water model and thus optimized both, the protein and the water interactions. In contrast, C36m used the Tip3p water model and optimized only protein interactions. In many cases, C36m+Tip3p underestimates radii of gyration compared to FRET or SAXS experiments. Such overly compact structural ensembles are believed to arise from an imbalance between protein-protein, protein-water, and water-water interactions, which might be due to Tip3p inaccuracies. Here, we aim at re-balancing these interactions by combining C36m with the Optimal Point Charge (OPC) water model. Recently, this C36m+OPC combination showed improved accuracy for the disordered domain of the measles virus nucleoprotein. Here we present a systematic assessment, comparing C36m+OPC, C36m+Tip3p, C36m+Tip4p, C22*, A03ws, A99sb-ws, and A99disp for five IDPs, as well as a subset of those for five globular proteins, a set of disordered AGQ-repeat peptides, and the fast folding miniprotein CLN025. We compared extensive MD simulations (> 8.5 ms) with SAXS, NMR, circular dichroism, photo-induced electron transfer (PET), T-jump infrared spectroscopy, and X-ray crystallography measurements. We found that combining C36m with OPC improved accuracy over C36m+Tip3p for IDP ensembles without compromising its accuracy for globular proteins. While also the kinetics of the AGQ-peptides were more accurate for C36m+OPC, those of CLN025 folding were less accurate. Overall, C36m+OPC showed similar accuracy as A99sb-ws and A99disp, the latter is currently considered among the most accurate protein force fields.

biophysics↗