Canary in the cardiac-valve coal mine: Flow velocity and inferred shear during prosthetic valve closure --predictors of blood damage and clotting

Abstract

ObjectiveTo demonstrate a clear link between predicted blood shear forces during valve closure and thrombogenicity that explains the thrombogenic difference between tissue and mechanical valves and provides a practical metric to develop and refine prosthetic valve designs for reduced thrombogenicity. MethodsPulsatile and quasi-steady flow systems were used for testing. The time-variation of projected open area (POA) was measured using analog opto-electronics calibrated to projected reference orifice areas. Flow velocity determined over the cardiac cycle equates to instantaneous volumetric flow rate divided by POA. For the closed valve interval, data from quasi-steady back pressure/flow tests was obtained. Performance ranked by derived maximum negative and positive closing flow velocities, evidence potential clinical thrombogenicity via inferred velocity gradients (shear). Clinical, prototype and control valves were tested. ResultsBlood shear and clot potential from multiple test datasets guided empirical optimization and comparison of valve designs. Assessment of a 3-D printed prototype valve design (BV3D) purposed for early soft closure demonstrates potential for reduced thrombogenic potential. ConclusionsThe relationship between leaflet geometry, flow velocity and predicted shear at valve closure illuminated an important source of prosthetic valve thrombogenicity. With an appreciation for this relationship and based on our experiment generated comparative data, we achieved optimization of valve prototypes with potential for reduced thrombogenicity. Competing InterestsNone declared. Financial DisclosureThis research has been done on a pro bono basis by all authors. Graphical AbstractVisualization of water jetting through closed mechanical heart valve under steady flow. Under pulsatile conditions, similar jet patterns near valve closure and leaflet rebound are likely. Dynamic metrics for several valves assessed in vitro are important in prediction of comparable blood cell damage and potential life-threatening thrombotic outcomes. Red star indicates moment of valve closure. O_FIG O_LINKSMALLFIG WIDTH=200 HEIGHT=169 SRC="FIGDIR/small/497372v11_ufig1.gif" ALT="Figure 1"> View larger version (57K): org.highwire.dtl.DTLVardef@95677forg.highwire.dtl.DTLVardef@bdc94dorg.highwire.dtl.DTLVardef@1eaf113org.highwire.dtl.DTLVardef@1792ade_HPS_FORMAT_FIGEXP M_FIG C_FIG CENTRAL MESSAGEA derived laboratory metric for valve closing flow velocity offers a way to rank valve models for potential blood damage. These results provide new insight and a mechanistic explanation for prior clinical observations where aortic and mitral valve replacements differ in thrombogenic potential and anticoagulation requirement. The study suggests a path forward to design and evaluate novel mechanical valve models for future development. As multiple modifications to mechanical and bioprosthetic valves have not resolved chronic shortcomings related to thrombogenicity and durability, a new development avenue was required to lead to eliminate thrombogenicity in the former and extend durability in the latter. PERSPECTIVEProsthetic mechanical valve devices cause blood cell damage. Activation of the coagulation cascade is initiated by dynamic valve function. Design innovation focusing on valve closure behavior may reduce valve thrombogenic potential. Our study demonstrates that valve design can be empirically optimized with emphasis on that phase. SIGNIFICANCEEmphasis on open valve performance has encouraged a long-standing bias while under appreciation of the closing phase vital to identification of potential thrombogenic complications persist. Our multiple data sets are useful in challenging this bias. Dynamic motion(s) of mechanical valves and derived regional flow velocity are impacted by valve geometry. Focus on valve closure dynamics may lead to the development of potentially less thrombogenic prototype valves. Laboratory experiments support the supposition that valve regional flow velocity is associated with valve thrombogenic potential. This study compares three clinical valves and two experimental prototypes.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Scotten, L. N., Siegel, R., Blundon, D. J., Deutsch, M.-A., Martin, T. R. P., Dutton, J. W., Kolahdouz, E. M., Griffith, B. E.. 2022-06-26. Canary in the cardiac-valve coal mine: Flow velocity and inferred shear during prosthetic valve closure --predictors of blood damage and clotting. https://doi.org/10.1101/2022.06.23.497372

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related preprints

Engineered microvessel networks -- Longitudinal transcriptomic profiling validates chip-grown tissues as a robust model for studying vasculature

Vascularised Organ-on-Chip (OoC) models are promising New Approach Methodologies (NAMs), but their translation requires validation for defined Contexts of Use (CoU). Here, we combine microfluidic microvessel networks (MVNs) with longitudinal bulk and single-cell RNA sequencing to determine how cell provenance influences engineered vascular phenotype. Stromal cell identity produced distinct network architectures and transcriptional trajectories, demonstrating that support cells, commonly treated as interchangeable, represent an important CoU variable. Primary human pancreatic microvascular endothelial cells failed to form stable perfusable networks, but supplementation with 10% Human Umbilical Vein Endothelial Cells (HUVECs) or Blood Outgrowth Endothelial Cells (BOECs) rescued self-assembly. Single-cell transcriptomics identified specialised endothelial populations associated with vascular morphogenesis, including tip, stalk, phalanx, endothelial to mesenchymal transition (EndMT) and pericyte-like endothelial cells. Failed networks lacked coherent endothelial specialisation, whereas rescued networks restored stabilising cell states and revealed FLRT2--UNC5B signalling within pericyte-like HUVECs, recapitulating a pathway implicated in angiogenic development in vivo. These findings establish cell provenance as a critical determinant of microvascular NAM performance and demonstrate how transcriptomic validation can define and strengthen the biological relevance of microvessel-on-chip models for specific CoUs.

bioengineering↗

Benchmarking foundation models for resting-state functional magnetic resonance imaging time series in cognitive aging research

Foundation models trained on large neuroimaging datasets could help smaller studies investigate cognition from brain recordings. However, whether the foundation model-derived representations improve assessment of current cognition, prediction of follow-up cognition, or characterization of cognitive change remains unclear. We evaluated nine configurations from five model families that process regional resting-state functional magnetic resonance imaging time series, using 3,103 scans from 1,653 participants across two observational and five intervention aging cohorts. Primary frozen representations showed no consistent superiority over conventional functional connectivity for concurrent cognitive associations or diagnostic classification. Changing the readout improved selected concurrent cognitive associations without updating model weights, indicating that performance depended partly on how representations were extracted. Additional post-training and embedding alignment yielded gains in selected settings but did not consistently improve performance across outcomes. In the primary longitudinal analyses, imaging representations did not establish added value beyond baseline clinical measures. These findings show that readout choice influences concurrent cognitive associations, while the added value of these representations for longitudinal outcomes remains unestablished.

bioengineering↗

PROST: An Open Toolbox, Harmonised Multi-Lab Dataset, and Meta-Analysis of Ankle-Foot Roll-Over Shape

The ankle-foot roll-over shape (ROS) is the sagittal-plane path of the centre of pressure (CoP) during the stance phase of walking, expressed in a shank-based coordinate system. It describes the effective rocker geometry that emerges from both the shape and the elastic deformation of the ankle-foot system during walking, and informs prosthetic and orthotic design. Previous studies characterised ROS in small samples with differing laboratory protocols, so they have not systematically assessed inter-individual variability or its relation to anthropometrics. We developed the open-source Public ROS Toolbox (PROST) for standardised ROS computation, and used it to harmonise marker trajectories, ground reaction forces and CoP data from four public gait datasets collected in independent laboratories. For each leg, we fitted a circular arc to the pooled ROS of all its stance phases. In a one-stage individual-participant data meta-analysis, linear mixed-effects models, with dataset as a fixed effect and participant as a random intercept, related the arc radius and the horizontal coordinate of the circle centre to anthropometrics, and, in the two self-selected-speed datasets, to walking speed. The harmonised dataset comprises 249 participants and 5,667 stance phases at walking speeds of 0.7-8.5 km/h. Circular arcs fitted the ROS closely (mean RMS radial residual 1.9 mm). The radius had a median of 410.3 mm (interquartile range, IQR, 237.6 mm), and the circle centre lay posterior to the ankle (median -20.6 mm, IQR 49.5 mm). Between-participant differences, quantified by the intraclass correlation coefficient (ICC), accounted for 42% of the variance in log radius (95% CI 25-57%) and 47% of the variance in the horizontal coordinate of the circle centre (95% CI 24-65%). Anthropometrics explained a negligible share of this variance (radius: pseudo-R2 -0.001, 95% CI -0.03 to 0.12), and height normalisation did not reduce radius variability (robust coefficient of variation 43% with and without normalisation). Faster self-selected walking was associated with a larger radius and a more posterior circle centre. To our knowledge, this study provides the first harmonised multi-lab ROS dataset. Together with the open-source toolbox, it is publicly available to support reproducible research in gait biomechanics, rehabilitation device design and clinical gait assessment. ROS geometry was participant-specific but was not predicted by age, sex, height or mass. A device ROS scaled from anthropometrics is therefore unlikely to match an individual's ROS, motivating studies of whether matching it, through device geometry and stiffness, improves gait mechanics.

bioengineering↗