bioRxiv · 10.1101/2020.11.02.365940
Elemental Mapping of Labeled Biological Specimens at Intermediate Energy Loss in an Energy-Filtered TEM acquired using a Direct Detection Device
Abstract
The multi-color or single-color EM that was developed previously, by the pseudo-colored overlay of the core-loss or high-loss EFTEM elemental map/s of the lanthanide onto the conventional image, the lanthanide chelates conjugated to diaminobenzidine being sequentially deposited as a result of selective oxidization by orthogonal photosensitizers / peroxidases. The synthesis of the new second generation lanthanide DABs, which contains 4 times more lanthanide per DAB, gives significant signal amplification and enabling collection of elemental maps at much lower energy-loss regions more favorable. Under the same experimental conditions, acquiring EFTEM elemental maps for the lanthanides at the lower energy-loss of N4,5 edge instead of the core-loss M4,5 edge, provides ~4x increase in signal-to-noise and ~2x increase in resolution. The higher signal at the N4,5 edge, also allows for more sophisticated technique of EFTEM spectrum Image for the acquisition of elemental maps with very high signal fidelity.
Source connections
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Ramachandra, R., Mackey, M. R., Hu, J., Peltier, S. T., Nguyen-Huu, X., Ellisman, M. H., Adams, S. R.. 2020-11-03. Elemental Mapping of Labeled Biological Specimens at Intermediate Energy Loss in an Energy-Filtered TEM acquired using a Direct Detection Device. https://doi.org/10.1101/2020.11.02.365940