bioRxiv Science⌕ Search

Biology subjects

Holzhammer, T.

Publications and source records attributed to Holzhammer, T..

3 recordsLinked to original sources

Friend, not Foe: Lowered Tissue Reactivity to Long-Term Polyimide Implants

One of the biggest challenges for neurotechnology is the design of devices that are tolerated well by brain tissue, without sacrificing functionality and implantability. This study examined which design choices mitigate tissue damage and improve longevity, by varying probe features implanted in the cerebral cortex of mice. We report on a systematic, quantitative analysis of neuronal and inflammation markers across cortical depth. We implanted a total of 103 stiff silicon or flexible polyimide probes in 32 mice, varying their thicknesses and widths, which were either attached to the skull or not. A new, automated workflow to quantify immunohistochemical data examines: 1) the tissue loss caused by the implant, 2) the cortical neuronal density, and 3) the immune response expressed by astrocytic and microglial reaction. Flexible polyimide probes exhibited a clear advantage, with fewer lesions and weaker immune responses than stiff silicon probes. Furthermore, we observed a weaker influence of the shank cross-section. A cortical depth profile of immune reactivity revealed focal reactions at the device entry points in the superficial cortex and at the cortex-white matter boundary. This study gives important insights on optimizing device design parameters as well as surgical insights for improved tissue integration of intracortical electrode arrays.

bioengineering↗

Intracortical probe arrays with silicon backbone and microelectrodes on thin polyimide wings enable long-term stable recordings in vivo

ObjectiveRecording and stimulating neuronal activity across different brain regions requires interfacing at multiple sites using dedicated tools while tissue reactions at the recording sites often prevent their successful long-term application. This implies the technological challenge of developing complex probe geometries while keeping the overall footprint minimal, and of selecting materials compatible with neural tissue. While the potential of soft materials in reducing tissue response is uncontested, the implantation of these materials is often limited to reliably target neuronal structures across large brain volumes. ApproachWe report on the development of a new multi-electrode array exploiting the advantages of soft and stiff materials by combining 7-m-thin polyimide wings carrying platinum electrodes with a silicon backbone enabling a safe probe implantation. The probe fabrication applies microsystems technologies in combination with a temporal wafer fixation method for rear side processing, i.e., grinding and deep reactive ion etching, of slender probe shanks and electrode wings. The wing-type neural probes are chronically implanted into the entorhinal-hippocampal formation in the mouse for in vivo recordings of freely behaving animals. Main resultsProbes comprising the novel wing-type electrodes have been realized and characterized in view of their electrical performance and insertion capability. Chronic electrophysiological in vivo recordings of the entorhinal-hippocampal network in the mouse of up to 104 days demonstrated a stable yield of channels containing identifiable multi-unit and single-unit activity outperforming probes with electrodes residing on a Si backbone. SignificanceThe innovative fabrication process using a process compatible, temporary wafer bonding allowed to realize new Michigan style probe arrays. The wing-type probe design enables a precise probe insertion into brain tissue and long-term stable recordings of unit activity due to the application of a stable backbone and 7-m-thin probe wings provoking locally a minimal tissue response and protruding from the glial scare of the backbone.

neuroscience↗

Why not record from every channel with a CMOS scanning probe?

It is an uninformative truism to state that the brain operates at multiple spatial and temporal scales, each with each own set of emergent phenomena. More worthy of attention is the point that our current understanding of it cannot clearly indicate which of these phenomenological scales are the significant contributors to the brains function and primary output (i.e. behaviour). Apart from the sheer complexity of the problem, a major contributing factor to this state of affairs is the lack of instrumentation that can simultaneously address these multiple scales without causing function altering damages to the underlying tissue. One important facet of this problem is that standard neural recording devices normally require one output connection per electrode. This limits the number of electrodes that can fit along the thin shafts of implantable probes generating a limiting balance between density and spread. Sharing a single output connection between multiple electrodes relaxes this constraint and permits designs of ultra-high density probes. Here we report the design and in-vivo validation of such a device, a complementary metal-oxide-semiconductor (CMOS) scanning probe with 1344 electrodes; the outcome of the European research project NeuroSeeker. We show that this design targets both local and global spatial scales by allowing the simultaneous recording of more than 1000 neurons spanning 7 functional regions with a single shaft. The neurons show similar recording longevity and signal to noise ratio to passive probes of comparable size and no adverse effects in awake or anesthetized animals. Addressing the data management of this device we also present novel visualization and monitoring methods. Using the probe with freely moving animals we show how accessing a number of cortical and subcortical brain regions offers a novel perspective on how the brain operates around salient behavioural events. Finally, we compare this probe with lower density, non CMOS designs (which have to adhere to the one electrode per output line rule). We show that an increase in density results in capturing neural firing patterns, undetectable by lower density devices, which correlate to self-similar structures inherent in complex naturalistic behaviour. To help design electrode configurations for future, even higher density, CMOS probes, recordings from many different brain regions were obtained with an ultra-dense passive probe.

neuroscience↗